Measurement of Seebeck coefficient using a light pulse
C. Wood, D. Zoltan, G. Stapfer
Review of Scientific Instruments
Abstract
A high-temperature (1900 K) Seebeck coefficient apparatus is described in which small thermal gradients are generated in a sample by light pulses transmitted via light pipes. By employing an analog subtraction circuit, the Seebeck coefficient is displayed directly on an X–Y recorder. This technique presents a convenient, accurate, and rapid method for measuring the Seebeck coefficient in highly doped semiconductors as a function of temperature. The nature of the resulting display (X–Y recording) is a valuable tool in determining validity of the data. A straight line results (i.e., a minimum of hysteresis) only if all potential experimental errors are minimized. Under these conditions, the error of measurements of the Seebeck coefficient is estimated to be less than ±1%.
DOI: 10.1063/1.1138213